Instruction summary, Instruction descriptions, Instruction – INFICON RQCM - Quartz Crystal Microbalance Research System User Manual
Page 81: Summary -5, Descriptions -5

RQCM – RESEARCH QUARTZ CRYSTAL MICROBALANCE
COMPUTER INTERFACE
7-5
Length=2
Instruction Code
Receive code
Checksum
A value of 253 for the instruction byte indicates that this is a received status message. The
Instruction Code byte indicates the instruction code of the message that was received. The
following table shows a list of possible receive codes:
Receive Code
Description
0 Message
received
O.K.
1 Invalid
checksum.
2
Invalid instruction code.
3
Invalid message length.
4
Parameter(s) out of range.
5 Invalid
message.
7.11 INSTRUCTION SUMMARY
The following table is a list of valid instruction codes.
Instruction Code
Description
0 Send
RQCM
configuration
1
Initiate automatic data logging of binary values
2
Set analog to digital board configuration
3 Internal
command
4 Internal
command
5 Internal
command
6
Set Relay Outputs
7 Internal
command
8
Set interface address
7.12 INSTRUCTION DESCRIPTIONS
The following is a description of the valid instructions along with an example of how they are
used. All the examples assume the device address is 1.