RQCM – RESEARCH QUARTZ CRYSTAL MICROBALANCE

CRYSTALS, HOLDERS AND FLOW CELL

4-4

The minimum detectable mass change is typically a few ng/cm

2

and limited by the noise

specifications of the crystal oscillator and the resolution of the equipment used to measure

frequency shifts. For example, the INFICON RQCM has a frequency resolution of 0.03 Hz @ 6

MHz, therefore, its minimum detectable mass change is 0.37 ng/cm

2

.

The Sauerbrey equation relies on a sensitivity factor, C

f

, which is a fundamental property of the

QCM crystal. Thus, in theory, the QCM mass sensor does not require calibration. This ability to

calculate the mass sensitivity from first principles is obviously a very attractive feature of these

devices. However, it is very important to notice, that the Sauerbrey equation is only strictly

applicable to uniform, thin-film deposits originating from a low pressure (i.e. vacuum) gas

environment

21

. Thick deposits and operation in liquid environments or in contact with lossy films,

relies on the use of more complex equations relating the frequency shifts to mass loading, and

often requires calibration of the setup for accurate results. Several articles have been published on

simple ways to calibrate the mass sensitivity of QCMs for electrochemical applications

4

and for

vacuum thin-film deposition processes

5

6

7

, and some useful calibration guidelines are also

described herein.
Many studies have shown that the crystal's sensitivity is approximately Gaussian. The maximum

sensitivity is in the center of the crystal and it tapers off towards the edge of the active area

8

9

10

.

The mass sensitivity distribution has also been shown to become slightly more confined to the

electrode region as the mass loading is increased.

4.1.7 STABILITY

A sensor crystal cannot distinguish the difference between a frequency shift due to deposited

material or that due to other disturbances. Thus any extraneous factors, other than the deposited

mass, which may cause the quartz crystal to change its resonant frequency, must be properly

controlled. Factors that can influence the stability of a sensor crystal are categorized as follows

11

:

The crystal itself: Improper design, localized stress, damage to the crystal

The crystal holder: Improper seating of the crystal, large mechanical coupling between the

crystal and the holder

Thermal input: Radiation from evaporation source, radiation from substrate heater,

bombardment by charge particles, energy released by condensates

Stress: Thermal stress, stress release in the deposited materials

Temperature: See section 4.1.9 for data on frequency versus temperature for INFICON’s

crystals.

Other factors that can affect stability are humidity, shock, vibration and change in pressure.

Controlling those conditions is a must to insure accurate measurements of small mass changes

over long periods of time.

4.1.8

CRYSTAL LIFE EXPECTANCY

It is difficult to predict the useful life of a crystal since it depends on many factors. Some of these

factors are

11

:

The quality of the quartz

The amount of deposited material

The stress generated in the crystal due to deposited material