INFICON BCG450-SP ATM to Ultra-High Vacuum Triple Gauge (Profibus) User Manual

Page 18

Advertising
background image

18

tira41e1 (2005-06) BCG450SPv1.cp

If, in the Exception Status, one of bits 4 … 6 is set, the current warning can be read
in the parameter "Exception Detail Warning". The Exception Detail Warning pa-
rameter consists of a total of 13 bytes that inform on the error status of the gauge.

Due to the use of the expanded error output format, these bytes have the following
structure:

Byte

No

Name

Description

Value

0

Common Exception
Detail
Size

Indicates the number of subse-
quent bytes that contain the
Common Exception Detail
Warning.

2

1

Common Exception
Detail 0

Contains current error messages
from the Common Exception
Detail Warning group.

Depending on

warning status

2

Common Exception
Detail 1

Contains current error messages
from the Common Exception
Detail Warning group.

Depending on

warning status

3

Device Exception
Detail
Size

Indicates the number of subse-
quent bytes that contain the
Device Exception Detail Warning.

5

4

Device Exception
Detail 0

One of N Status Extension.

Depending on

warning status

5

Device Exception
Detail 1
(Capacitance
diaphragm sensor)

This error information from the
Common Exception Detail
Warning group refers to
Capacitance diaphragm sensor.

Depending on

warning status

6

Device Exception
Detail 2
(Capacitance
diaphragm sensor)

This error information from the
Common Exception Detail
Warnings group refers to
Capacitance diaphragm sensor.

Depending on

warning status

7

Device Exception
Detail 3
(Pirani)

This error information from the
Common Exception Detail Alarm
group refers to Pirani.

Depending on

warning status

8

Device Exception
Detail 4
(Pirani)

This error information from the
Common Exception Detail Alarm
group refers to Pirani.

Depending on

warning status

9

Device Exception
Detail 5
(BA)

This error information from the
Common Exception Detail Alarm
group refers to BA.

Depending on

warning status

10

Device Exception
Detail 6
(BA)

This error information from the
Common Exception Detail Alarm
group refers to BA.

Depending on

warning status

11

Manufacturer
Exception Detail
Size

Indicates the number of subse-
quent bytes that contain the
Device Exception Detail Warning.

1

12

Manufacturer
Exception Detail

Contains current error messages
from the Manufacturer Exception
Detail Alarm group.

Depending on

warning status



3.1.1.12 Exception Detail Warning

ID 28

Advertising