INFICON BCG450-SP ATM to Ultra-High Vacuum Triple Gauge (Profibus) User Manual

Page 29

Advertising
background image

tira41e1 (2005-06) BCG450SPv1.cp

29

Safe State

Coding

PV behavior

Zero

0

The Process Value (measured value ID 19) is set
to 0.

Full Scale

1

The Process Value (measured value ID 19) is set
to the full scale value (ID 24).

Hold Last Value

2

The Process Value is set to the last valid value
obtained in the EXECUTING status.

Use Safe Value

3

The Process Value (measured value ID 19) is set
to the Safe Value (ID 40).

The Safe Value is the value output with the Process Value Parameter (ID 19) when
an error occurs or the gauge goes to the NOT EXECUTING status. If this value is
set to zero, it will remain on zero when the data unit is changed.

Overrange is the highest valid measured value at which Reading Valid is still
on "1".


Pressure unit

Overrange

COUNTS

21602

Torr

1.5×10

-2

mTorr (Micron)

15

mbar

2×10

-2

Pascal

2

Underrange is the lowest valid measured value at which Reading Valid is still
on "1".

Pressure unit

Underrange

COUNTS

6397.95

Torr

3.7503×10

-10

mTorr (Micron)

3.7503×10

-7

mbar

5.0×10

-10

Pascal

5.0×10

-8



3.2.3.8 Safe Value

3.2.3.9 Overrange

3.2.3.10 Underrange

Advertising