INFICON BCG450-SP ATM to Ultra-High Vacuum Triple Gauge (Profibus) User Manual

Page 26

Advertising
background image

26

tira41e1 (2005-06) BCG450SPv1.cp

When the gauge is not in the EXECUTING status (ID 25, Device Block) or if there
is a device error, a value defined by Safe State is output as pressure value. You
can select among:

• "0"
• Full scale
• Last valid value
• Safe Value (user-definable in ID 40)

Safe State

Coding

PV behavior

Zero

0

The Process Value (measured value ID 19) is set
to 0.

Full Scale

1

The Process Value (measured value ID 19) is set
to the full scale value (ID 24).

Hold Last Value

2

The Process Value is set to the last valid value
obtained in the EXECUTING status.

Use Safe Value

3

The Process Value (measured value ID 19) is set
to the Safe Value (ID 40).

The Safe Value is the value output with the Process Value parameter (ID 19) when
an error occurs or the gauge goes to the NOT EXECUTING status. If this value is
set to zero, it will remain on zero when the data unit is changed.

Overrange is the highest valid measured value at which Reading Valid is still
on "1".

Pressure unit

Overrange

COUNTS

31000

Torr

750.06168

mTorr (Micron)

750061.68

mbar

1000

Pascal

100000


Underrange is the lowest valid measured value at which Reading Valid is still
on "1".

Pressure unit

Underrange

COUNTS

20480.7254

Torr

4.12534×10

-3

mTorr (Micron)

4.12534

mbar

5.5×10

-3

Pascal

0.55


3.2.2.7 Safe State

3.2.2.8 Safe Value

3.2.2.9 Overrange

3.2.2.10 Underrange