3 transducer block – INFICON BCG450-SP ATM to Ultra-High Vacuum Triple Gauge (Profibus) User Manual
Page 36
36
tira41e1 (2005-06) BCG450SPv1.cp
If the Setpoint Function is set to Atmosphere Control and the relay is closed, this
byte is set to one, if the relay is open or not energized, the value is zero.
Additionally the following information is copied into the Manufacturer Exeption
Detail Warning:
• Bit 1 of Manufacturer Exeption Detail Warning is set if relay A is energized and
atmospheric pressure is reached.
• Bit 2 of Manufacturer Exeption Detail Warning is set if relay B is energized and
atmospheric pressure is reached.
ID
Name
Structure Data type
Bytes
Access
Store
120
One Of N Status
Extension
Simple
UINT8
1
1_R
2_R
V
This parameter indicates whether the overrange or underrange of the gauge is
exceeded.
Bit7
Bit6
Bit5
Bit4
Bit3
Bit2
Bit1
Bit0
0
0
0
0
0
Underrange
Exceeded
Overrange
Exceeded
Reading
Invalid
If the gauge is operated in its overrange or underrange (p < 5×10
-10
mbar or
p > 1500 mbar), the corresponding bit and additionally the bit "Reading Invalid" is
set.
If an error occurs, the bit "Reading Invalid" as well as the corresponding error bits
in Device Block (ID 26, 27, 28 Device Block) are set.
ID
Name
Structure Data type
Bytes
Access
Store
101
Block Type
Simple
Octet string
4
2_R
102
Status Extension
Simple
UINT8
1
1/2_R
V
103
Sensor Alarm
Struct
Array of 2 bytes
2
1/2_R
V
104
Sensor Warning
Struct
Array of 2 bytes
2
1/2_R
V
According to the table in Appendix A, the Block Type ID has the value "13".
This parameter indicates whether the overrange or underrange of the Heat
Transfer Vacuum Gauge device instance is exceeded.
Bit7
Bit6
Bit5
Bit4
Bit3
Bit2
Bit1
Bit0
0
0
0
0
0
Underrange
Exceeded
Overrange
Exceeded
Reading
Invalid
If the instance is operated in its overrange or underrange, the corresponding bit
and additionally the bit "Reading Invalid" is set.
If an error occurs, the bit "Reading Invalid" as well as the corresponding error bits
in Device Block (ID 26, 27, 28 Device Block) are set.
This parameter indicates the detectable errors occurring in connection with the
Pirani measuring part. The present implementation allows detection of one error.
Bit7
Bit6
Bit5
Bit4
Bit3
Bit2
Bit1
Bit0
Byte 0
0
0
0
0
0
0
0
0
Byte 1
0
0
0
0
0
0
Electronics Failure
1)
0
1)
Electronics Failure includes a sensor error.
The Sensor Alarm bits defined here are copied into the Device Block ID 27 in the
"Device Exception Detail Alarm" range of the Pirani measuring part.
3.2.5.8 Atmosphere Reached
3.3 Transducer Block
3.3.1 One Of N Vacuum Gauge
Transducer Block /
SLOT 1
3.3.1.1 One Of N Status Extension
3.3.2 Heat Transfer Vacuum
Gauge Transducer
Block / SLOT 1 /
Instance 1
3.3.2.1 Block Type
3.3.2.2 Status Extension
3.3.2.3 Sensor Alarm