If the Setpoint Function is set to Atmosphere Control and the relay is closed, this
byte is set to one, if the relay is open or not energized, the value is zero.

Additionally the following information is copied into the Manufacturer Exeption
Detail Warning:

• Bit 1 of Manufacturer Exeption Detail Warning is set if relay A is energized and

atmospheric pressure is reached.

• Bit 2 of Manufacturer Exeption Detail Warning is set if relay B is energized and

atmospheric pressure is reached.


ID

Name

Structure Data type

Bytes

Access

Store

120

One Of N Status
Extension

Simple

UINT8

1

1_R
2_R

V

This parameter indicates whether the overrange or underrange of the gauge is
exceeded.

Bit7

Bit6

Bit5

Bit4

Bit3

Bit2

Bit1

Bit0

0

0

0

0

0

Underrange

Exceeded

Overrange

Exceeded

Reading

Invalid

If the gauge is operated in its overrange or underrange (p < 5×10

-10

mbar or

p > 1500 mbar), the corresponding bit and additionally the bit "Reading Invalid" is
set.

If an error occurs, the bit "Reading Invalid" as well as the corresponding error bits
in Device Block (ID 26, 27, 28 Device Block) are set.


ID

Name

Structure Data type

Bytes

Access

Store

101

Block Type

Simple

Octet string

4

2_R

102

Status Extension

Simple

UINT8

1

1/2_R

V

103

Sensor Alarm

Struct

Array of 2 bytes

2

1/2_R

V

104

Sensor Warning

Struct

Array of 2 bytes

2

1/2_R

V

According to the table in Appendix A, the Block Type ID has the value "13".

This parameter indicates whether the overrange or underrange of the Heat
Transfer Vacuum Gauge device instance is exceeded.

Bit7

Bit6

Bit5

Bit4

Bit3

Bit2

Bit1

Bit0

0

0

0

0

0

Underrange

Exceeded

Overrange

Exceeded

Reading

Invalid

If the instance is operated in its overrange or underrange, the corresponding bit
and additionally the bit "Reading Invalid" is set.

If an error occurs, the bit "Reading Invalid" as well as the corresponding error bits
in Device Block (ID 26, 27, 28 Device Block) are set.

This parameter indicates the detectable errors occurring in connection with the
Pirani measuring part. The present implementation allows detection of one error.

Bit7

Bit6

Bit5

Bit4

Bit3

Bit2

Bit1

Bit0

Byte 0

0

0

0

0

0

0

0

0

Byte 1

0

0

0

0

0

0

Electronics Failure

1)

0

1)

Electronics Failure includes a sensor error.

The Sensor Alarm bits defined here are copied into the Device Block ID 27 in the
"Device Exception Detail Alarm" range of the Pirani measuring part.

3.2.5.8 Atmosphere Reached

3.3 Transducer Block

3.3.1 One Of N Vacuum Gauge

Transducer Block /
SLOT 1

3.3.1.1 One Of N Status Extension

3.3.2 Heat Transfer Vacuum

Gauge Transducer
Block / SLOT 1 /
Instance 1

3.3.2.1 Block Type

3.3.2.2 Status Extension

3.3.2.3 Sensor Alarm

Advertising