Appendix c: electrical connections – INFICON BCG450-SP ATM to Ultra-High Vacuum Triple Gauge (Profibus) User Manual
Page 46
46
tira41e1 (2005-06) BCG450SPv1.cp
Appendix C: Electrical Connections
Technical data of BCG450 gauges
→ [1], [2], [3].
SP A
SP B
SP A
SP B
Degas
42 k
Ω
3
6
1
4
9
11
13
14
7
8
2
12
5
15
1.25 AT
24V
Degas
Ident.
10
-
-
-
1)
( )
+
-
Measuring
signal
-
2)
2)
8
9
1
15
D-Sub,15 pins,
female,
soldering side
Common (power GND 24V supply)
Ground (housing, vacuum connection)
Threshold values
Electrical connection
Pin 1
Relay switching function A, n.o. contact
2)
Pin 2
Measuring signal output
0 … +10.13 V
Pin 3
Threshold (setpoint) A
1)
0 … +10 V
Pin 4
Relay switching function A, com contact
2)
Pin 5
Supply common
0 V
Pin 6
Threshold (setpoint) B
1)
0 … +10 V
Pin 7
Degas on, active high
0 V/+24 V
Pin 8
Supply
+24 V
Pin 9
Relay switching function B, n.o. contact
2)
Pin 10
Gauge identification
Pin 11
Relay switching function B, com contact
2)
Pin 12
Measuring signal common
Pin 13
Do not connect
Pin 14
Do not connect
Pin 15
Do not connect
1)
Do not connect pin 3 and pin 6 for normal operation of the gauge. These pins
are reserved for adjustment of the setpoint potentiometers.
2)
Relay assignement can be reprogrammed (
→ 34).
Sensor cable connection