When the gauge is not in the EXECUTING status (ID 25, Device Block) or if there
is a device error, a value defined by Safe State is output as pressure value. You
can select among:

• "0"
• Full scale
• Last valid value
• Safe Value (user-definable in ID 40)

Safe State

Coding

PV behavior

Zero

0

The Process Value (measured value ID 19) is set
to 0.

Full Scale

1

The Process Value (measured value ID 19) is set
to the full scale value (ID 24).

Hold Last Value

2

The Process Value is set to the last valid value
obtained in the EXECUTING status.

Use Safe Value

3

The Process Value (measured value ID 19) is set
to the Safe Value (ID 40).

The Safe Value is the value output with the Process Value Parameter (ID 19) when
an error occurs or the gauge goes to the NOT EXECUTING status. If this value is
set to zero, it will remain on zero when the data unit is changed.

Overrange is the highest valid measured value at which Reading Valid is still
on "1".

Pressure unit

Overrange

COUNTS

31352

Torr

1125.09252

mTorr (Micron)

1125092

mbar

1500

Pascal

150000

Underrange is the lowest valid measured value at which Reading Valid is still
on "1".

Pressure unit

Underrange

COUNTS

25000

Torr

0.75006168

mTorr (Micron)

750.06168

mbar

1

Pascal

100

3.2.4.7 Safe State

3.2.4.8 Safe Value

3.2.4.9 Overrange

3.2.4.10 Underrange