INFICON BCG450-SP ATM to Ultra-High Vacuum Triple Gauge (Profibus) User Manual

Page 41

Advertising
background image

tira41e1 (2005-06) BCG450SPv1.cp

41

ID

Name

Structure Data type

Bytes

Access Store

14

Atmosphere
gauge Block State
Command

REC

2

1_W
2_W

101

Block Type

Simple

Octet string

4

2_R

The Block Type ID has the value: "01 00 00 00" (octet string).

This service is used for activating the adjustment of the Piezo sensor to the
Capacitance diaphragm sensor via Profibus.

The atmosphere pressure sensor can be adjusted to the internal (vacuum cham-
ber) Capacitance diaphragm sensor.

If you want to adjust the atmosphere pressure sensor to the Capacitance
diaphragm sensor, proceed as follows:

n

Vent the vacuum system to atmospheric pressure

o

Send the "Release User ATM adjust" command

p

Perform the "User ATM adjust command" as specified below.

After this procedure, the atmosphere pressure sensor will give the same pressure
value as the Capacitance diaphragm sensor at atmospheric pressure.

ID

Name

Structure Data type

Bytes

Access

14

Atmosphere gauge Block
State Command

Record

→ below

2

1/2_W

Transition Command

Simple

Unsigned8

1

Transition Command
Target Value

Simple

Usingned8

1

State

Command

Name

Description

0

Inactive

No action

128

Adjust
Piezo /
CDG

This service is used for activating the adjustment of
the piezo sensor to the Capacitance diaphragm sen-
sor.

The Transition Command Target Value can have the
following values:

0

⇒ executes a User ATM adjust

1

⇒ releases User ATM adjust

3.3.4 Atmosphere Sensor

Transducer Block (Piezo
Sensor) / SLOT 1 /
Instance 4

3.3.4.1 Block Type

3.3.4.2 Atmosphere Sensor Block

State Command