Setpoint function – INFICON BCG450-SP ATM to Ultra-High Vacuum Triple Gauge (Profibus) User Manual

Page 35

Advertising
background image

tira41e1 (2005-06) BCG450SPv1.cp

35

Atmospheric pressure

× N

Atmospheric pressure threshold

= —————————————

100

Where:

Atmospheric pres-
sure threshold

[mbar] :

If the pressure inside the vacuum chamber rises
above this threshold, the relay "Atmosheric pres-
sure reached" is energized.

Atmospheric
pressure

[mbar] :

Atmospheric pressure (100%, ambient)

1)

N

[%] :

"Percentage of Atmosphere"
(1 … 140%, default value = 5%)

(Example values in the diagram below are printed in italic)

M

ea

su

re

d

va

lue

Measuring signal

(Pressure p)

833 mbar

816 mbar

1500 mbar

980 mbar

Atmospheric pressure (100%, ambient)

"Percentage of Atmosphere"

(85%)

(Atmospheric pressure
threshold)

Hysteresis

(2% of atmospheric
pressure threshold)

Relay status "Atmospheric pressure reached"

Upper limit of measuring range of gauge

1)

1)

The (piezo) sensor measuring atmospheric pressure can be calibra-

ted to the value of the Capacitance Diaphragm Gauge pressure
(

→  41).

The value "Percentage of Atmosphere" has to be entered as follows:

N'

=

10 × N

Where:

N

=

"Percentage of Atmosphere", as described above

N'

=

Value to be entered in ID202
(range 10 … 1400

1 … 140%)

Example: For a "Percentage of Atmosphere" of 50(%) enter 500 in ID202.

Setpoint function
"

Atmosphere Control"

ID202 value