3 cyclic message telegrams – INFICON BPG402-SP ATM to Ultra-High Vacuum Gauge (Profibus) User Manual

Page 11

Advertising
background image

tira47e1 (2005-08) BPG402SPv1.cp

11

The message telegrams listed below are implemented in the gauge. They can be
selected according to requirements. When selecting a message telegram, ascertain
what output format of the measured value (integer/float) is required and whether a
parameter channel is needed or not. The gauge can also be operated in such a
way that the master does not transmit any output data to the slave.

Standard

telegram

Master

Slave

Byte

Meaning

1

M

⇒ S

0

Transition Command

1

Transition Command Value for Hot Ion Gauges

2

M

⇒ S

0 … 7

Parameter Channel

8

Transition Command

9

Transition Command Value for Hot Ion Gauges

3

M

⇒ S

0 … 7

Parameter Channel

4

S

⇒ M

0

Exception status

1

One Of N status extension

2

One Of N PV selector

3 … 4

Process value UINT16

5

S

⇒ M

0

Exception status

1

One Of N status extension

2

One Of N PV selector

3 … 6

Process value float

6

S

⇒ M

0 … 7

Parameter channel

8

Exception status

9

One Of N status extension

10

One Of N PV selector

11 … 12 Process value UINT16

7

S

⇒ M

0 … 7

Parameter channel

8

Exception status

9

One Of N status extension

10

One Of N PV selector

11 … 14 Process value float

Depending on the standard telegrams used, the respective configuration data have
to be transmitted to the gauge during the configuration phase (

→ table on  5).

2.3 Cyclic Message

Telegrams

Configuration data