INFICON BPG402-SP ATM to Ultra-High Vacuum Gauge (Profibus) User Manual

Page 19

Advertising
background image

tira47e1 (2005-08) BPG402SPv1.cp

19

Bit

Common Exception Detail 0

Bit

Common Exception Detail 1

0

0

0

0

1

0

1

0

2

EPROM exception

2

0

3

EEPROM exception

3

0

4

RAM exception

4

0

5

0

5

0

6

0

6

0

7

0

7

0

The warning bits are set in the same way as the error bits because here, warnings
have the same meaning as errors.

Bit

Device Exception Detail 0

0

Bit set if

Reading Invalid

1

Bit set if

Device Overrange

2

Bit set if

Device Underrange

3

0

4

0

5

0

6

0

7

0

This byte is a copy of the One Of N

Vacuum Gauge Transducer Block.

Bit

Device Exception Detail 1

Referring to Pirani

Bit

Device Exception Detail 2

Referring to Pirani

0

0

0

0

1

0

1

Electronics / sensor warning

2

0

2

0

3

0

3

0

4

0

4

0

5

0

5

0

6

0

6

0

7

0

7

0

This byte is a copy of Sensor

Warning byte 0 of the Pirani

Transducer Block.

This byte is a copy of Sensor

Warning byte 1 of the Pirani

Transducer Block.

Bit

Device Exception Detail 3
Referring to BA

Bit

Device Exception Detail 4
Referring to BA

0

Sensor filament 1 Warning

0

0

1

Sensor filament 2 Warning

1

Electronics / sensor warning

2

0

2

0

3

0

3

Pressure too high for degas

4

0

4

0

5

0

5

0

6

0

6

0

7

0

7

0

This byte is a copy of Sensor

Warning byte 0 of the BA Transducer

Block.

This byte is a copy of Sensor

Warning byte 0 of the BA Transducer

Block.

Common Exception
Detail Warning

Device Exception
Detail Warning

Advertising