INFICON BPG402-SP ATM to Ultra-High Vacuum Gauge (Profibus) User Manual

Page 36

Advertising
background image

36

tira47e1 (2005-08) BPG402SPv1.cp

This service is used for activating degas and emission modes via Profibus.

ID

Name

Structure

Data type

Bytes Access

Range

14

Hot Cathode
Block State
Command

Rec

→ below

2

1/2_W

Byte 0

Transition
Command

Simple

UINT8

1

→ below

Byte 1

Value

Simple

UINT8

1

→ below

Transition

Command

Name

Description

0

Inactive

No action

1

Set Degas
State

Activates/deactivates the Degas mode

This service is used for activating/deactivating the
Degas mode (p <7.2

×10

-6

mbar).

If the Degas mode has not been deactivated with
Degas Off before, it is automatically turned off after
3 minutes.

The Degas State can only be changed if
p <7.2

×10

-6

mbar, otherwise the command returns the

error "Object State Conflict".

The Transition Command Data Field Value can have
the following values:
0

⇒ Degas Off

1

⇒ Degas On

3

Set
Emission
State

Turns the emission On or Off according to the target
value

0

⇒ Switch emission Off

1

⇒ Switch emission On

128

Emission
User Mode

The device supports two modes:

Automatic Mode:

Emission is switched on or off by the integrated Pirani
automatically.

If the emission is switched on and is then switched off
manually, the emission will be switched on again only
if the pressure rises above the 3.2

×10

-2

mbar and then

drops below 2.4

×10

-2

mbar.

Manual Mode:

Emission is switched on or off by the command
"Set Emission State"
• If during run time the pressure exceeds 3.2×10

-2

mbar, the emission is switched off automatically.
The measurement value equals the Pirani value.

• If the emission is switched off by the user while the

pressure is still in BA range, the measurement
value equals the Pirani value. There is no auto-
matic change to the BA range by the BPG402.

0

⇒ sets Manual Mode

1

⇒ sets Automatic Mode

3.3.3.12 Hot Cathode Block State

Command ID 14