INFICON BPG402-SP ATM to Ultra-High Vacuum Gauge (Profibus) User Manual

Page 27

Advertising
background image

tira47e1 (2005-08) BPG402SPv1.cp

27

Instance 2 of the Analog Sensor Input Function Block describes the functionality of
the BA measuring part of the gauge.

In Instance 2, the same attributes as in Instance 1 are supported:

ID

Name

Structure Data type

Bytes

Access

Store

19

Process Value
(PV)

Simple

According to
Data Type value
(parameter 21)

-

0_XI
1/2_R

V

20

Status

Simple

UINT8

1

0_XI
1/2_R

V

21

Data Type

Simple

UINT8

1

2_R/W

N

22

Data Units

Simple

UINT16

2

2_R/W

N

23

Reading Valid

Simple

Boolean

1

1/2_R

V

24

Full Scale

Simple

According to
Data Type value
(parameter 21)

-

1/2_R

N

39

Safe State

Simple

UINT8

1

1/2_R/W

N

40

Safe Value

Simple

According to
Data Type value
(parameter 21)

-

1/2_R/W

N

44

Overrange

Simple

According to
Data Type value
(parameter 21)

-

1/2_R

N

45

Underrange

Simple

According to
Data Type value
(parameter 21)

-

1/2_R

N

The Process Value contains the measured value of the BA Device Instance in the
currently selected data unit (ID 22) and in the selected data type (ID 21).

If the device is not in the EXECUTING status (ID 25, Device Block), the value
defined in the Safe State is output.

Values output in the data unit COUNTS can be converted into a pressure value by
means of the following formulas:

Calculation of the pressure (PV is the abbreviation used for Process Value):

PV

mbar

=

10

(COUNTS / 2000) - 12.5

PV

Torr

=

0.75006168 × PV

mbar

PV

Micron

=

10

-3

× PV

Torr

PV

Pa

=

100 × PV

mbar

This parameter is remains on "0".

Two data types are supported: Float and Integer16.

In cyclic data traffic, the data type cannot be modified. After the gauge is switched
on, the value stored in the EEPROM is loaded. The data type setting can only be
modified when the gauge is in the IDLE status. By defining the configuration data
for cyclic data traffic (selection of standard telegram), the data type used in the
selected standard telegram is taken over. All settings previously made in acyclic
data traffic are thus overwritten.

3.2.3 Analog Sensor Input

Function Block
SLOT 1 / Instance 2

3.2.3.1 Process Value ID 19

3.2.3.2 Status ID 20

3.2.3.3 Data Type ID 21