INFICON BPG402-SP ATM to Ultra-High Vacuum Gauge (Profibus) User Manual

Page 35

Advertising
background image

tira47e1 (2005-08) BPG402SPv1.cp

35

This parameter indicates the value of the emission current in mA.

Emission currents [mA]

0

0.025

5

20

The Parameter defines / indicates the active filament (

→ Mode Filament Selection).

Active Filament

Byte value

Meaning

1

1

filament 1 active

2

2

filament 2 active

This parameter indicates whether the gauge is in Degas mode.

Degas Status

Meaning

FALSE

Degas Off

TRUE

Degas On

For the BPG402-SP, the value of the parameter is the same as the parameter
value of Active Filament (ID 108).

The parameter defines the operation mode of the filaments:

Mode Filament

Selection

(Filament User

Mode)

Byte

value

Meaning

Auto

0

Automatic Mode (default):
The gauge alternates between filaments every
time the emission is switched on.

Manual

1

Manual Mode:
The gauge uses the filament selected by the
attribute Active Filament (ID 108).

This parameter indicates whether the gauge is in Automatic or Manual Emission
Mode. The mode is set by Transition command 18 (ID 14). Automatic Mode is
selected by default.

Emission User

Mode State

Byte

value

Meaning

FALSE

0

Manual Mode

TRUE

1

Automatic Mode



3.3.3.6 Emission Current ID 106

3.3.3.7 Active Filament ID 108

3.3.3.8 Degas Status ID 109

3.3.3.9 Active Degas Filament

ID 116

3.3.3.10 Mode Filament Selection

ID 119

3.3.3.11 Emission User Mode

State ID 201