3 transducer block, 1 one of n vacuum gauge transducer block / slot 1 – INFICON BPG402-SP ATM to Ultra-High Vacuum Gauge (Profibus) User Manual

Page 32

Advertising
background image

32

tira47e1 (2005-08) BPG402SPv1.cp


ID

Name

Structure Data type

Bytes

Access

Store

120

One Of N Status
Extension

Simple

UINT8

1

1/2_R

V

This parameter indicates whether the overrange or underrange of the gauge is
exceeded.

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

0

0

0

0

0

Underrange

Exceeded

Overrange

Exceeded

Reading

Invalid

If the gauge is operated in its overrange or underrange (p < 5×10

-10

mbar or

p >1000 mbar), the corresponding bit and additionally the bit "Reading Invalid" is
set.

If an error occurs, the bit "Reading Invalid" as well as the corresponding error bits
in Device Block (ID 26, 27, 28 Device Block) are set.



ID

Name

Structure Data type

Bytes

Access

Store

101

Block Type

Simple

Octet string

4

2_R

N

102

Status Extension

Simple

UINT8

1

1/2_R

V

103

Sensor Alarm

Struct

Array of 2 bytes

2

1/2_R

V

104

Sensor Warning

Struct

Array of 2 bytes

2

1/2_R

V

According to the table in Appendix A, the Block Type ID has the value "13".

This parameter indicates whether the overrange or underrange of the Heat
Transfer Vacuum Gauge device instance (Pirani) is exceeded.

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Reserved

Underrange
Exceeded

1)

Overrange

Exceeded

2)

Reading
Invalid

3)

1)

Underrange: The parameter is set to "1", if the pressure value is below the
value defined in ID 45 (Analog Input Function Block Instance 1).

2)

Overrange: The parameter is set to "1", if the pressure value is above the value
defined in ID 44 (Analog Input Function Block Instance 1).

3)

Reading Invalid: The parameter is set to "1", if "Reading Valid" ID 23 (Analog
Input Function Block Instance 1) is set to"0".

If the instance is operated in its overrange or underrange, the corresponding bit
and additionally the bit "Reading Invalid" is set.

If an error occurs, the bit "Reading Invalid" as well as the corresponding error bits
in Device Block (ID 26, 27, 28 Device Block) are set.

This parameter indicates the detectable errors occurring in connection with the
Pirani measuring part.

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Byte 0

0

0

0

0

0

0

0

0

Byte 1

0

0

0

0

0

0

Electronics Failure

1)

0

1)

Electronics Failure includes a sensor error.

The Sensor Alarm bits defined here are copied into the Device Block ID 27 in the
"Device Exception Detail Alarm" range of the Pirani measuring part.

3.3 Transducer Block

3.3.1 One Of N Vacuum Gauge

Transducer Block /
SLOT 1

3.3.1.1 One Of N Status Extension

3.3.2 Heat Transfer Vacuum

Gauge Transducer
Block / SLOT 1 /
Instance 1

3.3.2.1 Block Type ID 101

3.3.2.2 Status Extension ID 102

3.3.2.3 Sensor Alarm ID 103