INFICON SKY CDGxxxD (DeviceNet) User Manual

Page 20

Advertising
background image

20

tira51e1 (2009-01) CDGxxxD.cp

Following is the instance 1 with the subclass extension of the capacitance mano-
meter (diaphragm gauge) part of the gauge. Many different types of gauges are
available. The behavior of all gauges is the same, but the fullscale value is
different.

Attribute

ID

Access rule

NV/

V

Name

DeviceNet data
type

Description of
attribute

Semantics of values

3 Set

/Cond.

→ below

NV Data type

USINT

Determines the
data type of value
and all related
attributes as speci-
fied in this table.

→ "Semantics" section below.
Int

C3

h

[default]

float

CA

h

4 Set

→ below

NV Data units

UINT

Determines the
units context of
value and all re-
lated attributes.

→ "Semantics" section below.
Supported Values:
Counts

1001

h

[default]

mbar

1308

h

Torr

1301

h

Pascal

1309

h

5

Get

V Reading valid

BOOL

Indicates that the
value attribute con-
tains a valid value.

0 = invalid
1 = valid
(invalid: e.g., not warmed up
yet)

6

Get

V Value

INT or specified
by data type

Analog input value The corrected, converted, cali-

brated final value of the sensor.
→ "Semantics" section below.

7

Get

V Status

BYTE

Alarm and warning
state of this object
instance.

→ "Semantics" section below.

8

Set

NV Alarm enable

BOOL

Enables the setting
of the alarm status
bits

0 = disable [default]
1 = enable

9

Set

NV Warning enable

BOOL

Enables the setting
of the warning
status bits

0 = disable [default]
1 = enable

10

Get

NV Full Scale

INT or specified
by data type

The value of full
scale for the sen-
sor.

The value of attribute value
corresponding to the full scale
calibrated measurement of the
sensor.
[default] = maximum allowable
value for the data type

11

Get

NV Offset-A data type

USINT

Determines the
data type of
attribute offset-A

[default] = INT (C3

h

)

12 Set

NV

Offset-A Specified

by

attribute 11

An amount added
prior to gain to
derive value

This value could be read at the
measurement device itself after
execution of zero adjust service.
0 = [default]

17

Set

NV Alarm trip point

high

Determines

the

value above which
an alarm condition
will occur

[default] = Maximum value for
its data type.

18

Set

NV Alarm trip point low INT or specified

by data type

Determines the
value below which
an alarm condition
will occur

[default] = Minimum value for its
data type.

19

Set

NV Alarm hysteresis

INT or specified
by data type

Determines the
amount by which
the value must
recover to clear an
alarm condition

[default] = 0

4.4.3.1 Instance Attributes of

Instance 1

Advertising