INFICON SKY CDGxxxD (DeviceNet) User Manual

Page 23

Advertising
background image

tira51e1 (2009-01) CDGxxxD.cp

23

16 Bits are used as sensor warnings. Sensor warning byte 0 is mapped to the
device exception detail warning 1, sensor warning byte 1 is mapped to the device
exception detail warning 2.

Data component

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Sensor warning byte 0

0 0 0 0 0 0 0

Not

at

temperature

Sensor warning byte 1

High

warning

exception

Low

warning

exception

0 0 0 0

Electronics

warning

0



8 Bits providing the current sensor alarm state of the instance. They are mapped to
device exception detail warning 0.

Data component

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Sensor alarm byte 0

0 0 0 0 0

Underrange

exceeded

Overrange

exceeded

Reading
invalid

1)

1)

Logical inversion of reading valid


The S-Analog Sensor Object provides the following common services:

Service

Code

Service Name

Description of Service

0E

h

Get_Attribute_Single

Returns the contents of the specified attribute.

10

h

Set_Attribute_Single

Modifies an attribute value.


The S-Analog Sensor Object provides the following common services:

Service

Code

Service Name

Description of Service

4B

h

Zero_Adjust

Used to calibrate the zero pressure value.
2 = zero adjust

The Zero_Adjust service is used to cause the S-Analog sensor object device to
modify its Offset-A attribute value (instance 1) based upon manufacturer specific
algorithms. The target value specified in the service request represents the actual
parametric measurement that the physical sensor should be reporting at the time of
the request.
There are no state transitions associated with the invocation of this service. It is,
therefore, incumbent upon the user to establish the device into the desired con-
figuration prior to, and during, the execution of these services. This will generally
involve exposing the sensor to a known environment and treating the values read
during execution of the services accordingly.
A success service response indicates that the service was accepted and the
application process started.

To perform the Zero_Adjust Request pump down to a value 2 decades below the
minimum pressure range of the gauge and then start the Zero_Adjust Request
Service with a target value of "0".

Parameter

Data type

Description

Semantics of values

Zero_Adjust
parameter
or target
value

UINT or
FLOAT
depending on
data type

The target
value for the
zero calibration

The value to which the attribute
6, value, will be set (always 0).
If not specified, the default value
of zero is used.


Sensor warning

Status extension

4.4.3.3 Common Services

4.4.3.4 Object–Specific

Services

Zero adjust algorithm

Zero_Adjust request service
data field parameters

Advertising