5 trip point object – INFICON SKY CDGxxxD (DeviceNet) User Manual

Page 27

Advertising
background image

tira51e1 (2009-01) CDGxxxD.cp

27

Class Code 35

h

= 53

d

The Trip Point Object models the action of trip points (two relays) of the CDG
(corresponding to physical outputs) (Discrete Output Object). A trip point value,
designated as a high or low trip point, is compared to the specified source value.
This trip point is intended to be used as a process control indicator.

Attribute ID Access

rule

Name

DeviceNet
data type

Description of attribute

1

Get

Revision

UINT

Revision of this object. Note: All class definitions
are required to include this class attribute.

2 Get Max

instance

UINT

Maximum instance number of an object currently
created in this class level of the device.





Attribute

ID

Access rule NV/

V

Name

DeviceNet
data type

Description of attribute

Semantics of values

5 Set

NV

Low

trip

point

INT or based
data type
attribute

Defines the value at or be-
low which a trip point con-
dition will occur.

→ "Semantics" section below
[default] = 0

6 Get

NV

Low

trip

enable

BOOL

Enables the low trip point
setting.

[default] = 0

7

Get

V Status

BOOL

State of this object
instance.

0 = trip point condition does not exist
(unasserted)
1 = trip point condition exists
(asserted)

10 Set

NV

Hysteresis

Same

as

high/low point
data type

Determines the amount by
which the input must re-
cover to clear a trip point
condition.

→ "Semantics" section
[default] = 0

12 Get

NV

Destination

Packed
EPATH

Specifies the path of the
destination attribute whose
value will be set by output.

Path to attribute 3 of Instance 1 or 2
of DOP Object.

13

Get

V Output

BOOL

Output of the object the
value of which is sent to
destination.

= Status of the relay

14 Get

NV

Source

Packed
EPATH

Specifies the path of the
source attribute whose
value is retrieved for Input.

Path to attribute 6 of instance 1 of S-
analog sensor object

15

Get

V Input

INT or speci-
fied by data
type

Input to the object whose
value is retrieved from
source.

Value of attribute 6 of instance 1 of S-
analog sensor object

16 Get

NV

Data

units

ENGUNITS

Units of input, trip point.

17 Get

NV

Data

type

USINT Data type of input.

[default] = INT

101 Set

NV

Percentage
of ATM

REAL

Multiplier for ATM control.

If high trip point enabled, this value
will be multiplied with pressure value
from S-Analog sensor instance 2
(ATM sensor) copied into trip point
high

102 Get

V

High

trip

point

INT or based
data type
attribute

Defines the value at or
above which a trip point
condition will occur.

[default] = 0

103 Get

NV

High

trip

enable

BOOL

Enables the high trip point
setting.

[default] = enabled

4.5 Trip Point Object

4.5.1 Class Attributes

4.5.2 Instance Attributes of

Trip Point Objects
Instance 1 & 2