Purpose, Technical data, General data – INFICON PEG100 Medium to Ultra-High Vacuum Gauge User Manual

Page 4: 2 technical data

Advertising
background image

1.1.1 Purpose

The PEG100 Penning Gauge is a compact active pres-
sure converter housing a measurement system as well
as the corresponding operating electronics. It has been
developed specifically for integration into vacuum
systems and offers a measurement range from 1·10

-9

to

1·10

-2

mbar.

The PEG100 is connected directly to the vacuum sys-
tem through its DN 25 KF or DN 40 CF flange.

The electrical connection is provided through a screened
8-way FCC 68 connector.

Moreover, the technical data as published in Section 1.2
must be observed.

1.2

Technical Data

1.2.1 General Data

Measurement range

1·10

-9

to 1·10

-2

mbar/Torr

Measurement uncertainty in the range
from 1·10

-8

to 1·10

-4

mbar:

Deviation from the characteristic

± 30 % of the displayed value

Average temperature coefficient of the output span

< 0.5 % / K of the displayed value

Reproducibility

< 4 % of the displayed value

Measurement principle

Cold cathode ionization according to Penning

Supply voltage

14.5 to 36 V DC, typ. 24 V DC

Ripple

2 V

pp

Power consumption

< 2 W

Protection

IP 40

Electromagnetic compatibility (EMI)

CE mark

interference tolerance to EN 50082-2, Tab. 1, 2, 3

interference emission levels to EN 50081-1 and

FCC Rules Part 15, Class B

Flammability

UL 94 - V2

4

tina14e1 (0106) PEG100.ga

1.1.1 Purpose

The PEG100 Penning Gauge is a compact active pres-
sure converter housing a measurement system as well
as the corresponding operating electronics. It has been
developed specifically for integration into vacuum
systems and offers a measurement range from 1·10

-9

to

1·10

-2

mbar.

The PEG100 is connected directly to the vacuum sys-
tem through its DN 25 KF or DN 40 CF flange.

The electrical connection is provided through a screened
8-way FCC 68 connector.

Moreover, the technical data as published in Section 1.2
must be observed.

1.2

Technical Data

1.2.1 General Data

Measurement range

1·10

-9

to 1·10

-2

mbar/Torr

Measurement uncertainty in the range
from 1·10

-8

to 1·10

-4

mbar:

Deviation from the characteristic

± 30 % of the displayed value

Average temperature coefficient of the output span

< 0.5 % / K of the displayed value

Reproducibility

< 4 % of the displayed value

Measurement principle

Cold cathode ionization according to Penning

Supply voltage

14.5 to 36 V DC, typ. 24 V DC

Ripple

2 V

pp

Power consumption

< 2 W

Protection

IP 40

Electromagnetic compatibility (EMI)

CE mark

interference tolerance to EN 50082-2, Tab. 1, 2, 3

interference emission levels to EN 50081-1 and

FCC Rules Part 15, Class B

Flammability

UL 94 - V2

4

tina14e1 (0106) PEG100.ga

Advertising