Measurement system, Signal output, Control inputs – INFICON PEG100 Medium to Ultra-High Vacuum Gauge User Manual

Page 5: Status output

Advertising
background image

Status displays

Operation (POWER)

orange LED

Ready to measure (ignited) READY

green LED

1.2.2 Measurement System

Measurement system

detachable

Vacuum connection

DN 25 KF or DN 40 CF

Degassing temperature

see Section 2.3.3

Dead volume

21 cm

3

approx.

Materials in contact with the medium

stainless steel, CrNi,

Al

2

O

3

ceramics, NiFe, Ni, Ti

Overpressure tolerance

10 bar abs.

(The limits for the flange connections must be observed)

Operating voltage

1.6 kV (current limited to < 0.5 mA)

Ignition voltage

2.8 kV (current limited to < 0.5 mA)

1.2.3 Signal Output

Signal output

0 to 10.6 V

Permissible load resistance

R

a

10 k

Measurement signal

0.66 V to 10 V;

logarithmic, 1.333 V per decade

Status signal (not ignited)

0.4 V

1.2.4 Control Inputs

Input resistance

R

E

: 10 k

approx.

High voltage cut-in with negative logic at Pin 7:

High voltage ON

at U < 2.5 V

High voltage OFF

at U > 4 V

or
High voltage cut-in with positive
logic at Pin 8:

High voltage ON

at U > 12 V

High voltage OFF

at U < 7 V

(for this also refer to Fig. 3 and Section 2.2.2).

1.2.5 Status Output

Ready to measure High level (13.5-35 V, max. 50 mA)
Error (not ignited, HV off)

0V

For this refer also to Section 2.2.4.

5

tina14e1 (0106) PEG100.ga

Status displays

Operation (POWER)

orange LED

Ready to measure (ignited) READY

green LED

1.2.2 Measurement System

Measurement system

detachable

Vacuum connection

DN 25 KF or DN 40 CF

Degassing temperature

see Section 2.3.3

Dead volume

21 cm

3

approx.

Materials in contact with the medium

stainless steel, CrNi,

Al

2

O

3

ceramics, NiFe, Ni, Ti

Overpressure tolerance

10 bar abs.

(The limits for the flange connections must be observed)

Operating voltage

1.6 kV (current limited to < 0.5 mA)

Ignition voltage

2.8 kV (current limited to < 0.5 mA)

1.2.3 Signal Output

Signal output

0 to 10.6 V

Permissible load resistance

R

a

10 k

Measurement signal

0.66 V to 10 V;

logarithmic, 1.333 V per decade

Status signal (not ignited)

0.4 V

1.2.4 Control Inputs

Input resistance

R

E

: 10 k

approx.

High voltage cut-in with negative logic at Pin 7:

High voltage ON

at U < 2.5 V

High voltage OFF

at U > 4 V

or
High voltage cut-in with positive
logic at Pin 8:

High voltage ON

at U > 12 V

High voltage OFF

at U < 7 V

(for this also refer to Fig. 3 and Section 2.2.2).

1.2.5 Status Output

Ready to measure High level (13.5-35 V, max. 50 mA)
Error (not ignited, HV off)

0V

For this refer also to Section 2.2.4.

5

tina14e1 (0106) PEG100.ga

Advertising