6 interfaces „interfce, Interfaces „interfce, 7 configuration – JUMO 701550 di 308 - Digital Indicator Operating Manual User Manual
Page 60

7 Configuration
60
7.6 Interfaces „INTERFCE“
The interface parameters for the RS422/485 or
PROFIBUS-DP interface have to be configured in
order to communicate with PCs, bus systems and
peripheral devices.
Configuration
Analog inputs
Limit
comparators
Outputs
Binary
functions
Display /
Operation
Interfaces
➔ RS422485 (Modbus) ➔
Parameter
Value/
Selection
Description
Protocol
P R O T O C O L
0
1
Modbus
Modbus integer
Baud rate
BAUD RATE
0
1
2
9600 bps
19200 bps
38400 bps
Data
format
DFORMAT
0
1
2
3
8 data bits, 1 stop bit, no parity
8 data bits, 1 stop bit, odd parity
8 data bits, 1 stop bit, even parity
8 data bits, 2 stop bits, no parity
Device address
ADDRESS
0…1…
255
Address in data network
Min. response
time
(Setup)
0…500
Time period in milli-seconds that elapses
between the request of a device in the data
network and the response of the display
instrument.
Adjustment only possible in the setup
program:
-> Interfaces -> RS422/RS485
Factory settings are shown bold.