Operation – Leica Biosystems EG1160 User Manual

Page 20

Advertising
background image

20

Instruction manual V 4.1 – 05/2001

SET TEMPERATURES
COLD PLAT

XX°C

SET TEMPERATURES
CASS BATH

XX°C

SET TEMPERATURES
MOLD TRAY

XX°C

SET TEMPERATURES
WORK SURF

XX°C

TODAY IS ...
MO

THE TIME IS ...

10:00

HRS

START WORK AT ...

10:00

HRS

STOP WORK AT ...

16:00

HRS

WORKDAYS
MO

YES

SET TEMPERATURES
PAR TANK

XX°C

etc. ...

SET Mode (Programming)

6. The SET mode is activated by pushing the SET/ACT button. The SET mode

is used for programming only. The programmable parameters are as fol-
lows:

-

Paraffin reservoir temperature

-

Cold plate temperature

-

Cassette bath temperature

-

Mold warmer temperature

-

Work area temperature

-

Actual day

-

Actual time

-

Planned start time

-

Planned finishing time

-

Workdays

The instrument is programmed to have all functional units ready to oper-
ate at the preselected time on the preselected workdays, i.e. all heated
and refrigerated components are turned on automatically so that they
will have reached the preselected temperatures when the operator starts
working (see chapter 4 "Technical data").

The MENU button is used to scroll through the individual items of the
menu.

Once the last item of the menu ("WORKDAYS ARE .. ") is reached, the first
item ("SET TEMPERATURES: PAR TANK") will be indicated again after
pressing the MENU button.

SET TEMPERATURES
PAR-TANK

XX°C

10.

Operation

Advertising