FUNCTIONAL UNIT

SYMPTOM

CAUSE

ACTION

Initial operation

Instrument does not turn

Main switch (on the back)

Turn on main switch

on

not in ON position

ON/OFF button pushed

Hold down ON/OFF

not long enough

button for some
seconds

Paraffin reservoir

Paraffin quantity level indic-

Paraffin not entirely

Wait until the wax is

ation does not correspond

molten

molten completely;

to the actual quantity

then check again.

Paraffin does not melt

Selected temperature

Increase temperature

too low?

of paraffin reservoir

Heater not activated?

Check if paraffin
reservoir heater is
activated

Temperature readout

Wrong temperature

Check temperature

incorrect

selected?

setting and

oractivation
no heating

Instrument failure

Technical service

Paraffin dispenser

No paraffin flow

Paraffin not entirely

Check the consistency

molten

of the wax in the reser
voir
and wait for some time
if necessary

Foot pedal not properly

Check foot pedal

connected

connection

Whenever a malfunction occurs, please ensure that it is not the result of an operating mistake.

The list below includes the most frequent operating errors.

Equipment malfunctions are indicated as error messages (error codes 1 to 13) in the LCD display. If such an error
code is displayed, please contact your local Leica service center.