Troubleshooting – Leica Biosystems EG1160 User Manual
Page 28
28
Instruction manual V 4.1 – 05/2001
FUNCTIONAL UNIT
SYMPTOM
CAUSE
ACTION
Initial operation
Instrument does not turn
Main switch (on the back)
Turn on main switch
on
not in ON position
ON/OFF button pushed
Hold down ON/OFF
not long enough
button for some
seconds
Paraffin reservoir
Paraffin quantity level indic-
Paraffin not entirely
Wait until the wax is
ation does not correspond
molten
molten completely;
to the actual quantity
then check again.
Paraffin does not melt
Selected temperature
Increase temperature
too low?
of paraffin reservoir
Heater not activated?
Check if paraffin
reservoir heater is
activated
Temperature readout
Wrong temperature
Check temperature
incorrect
selected?
setting and
oractivation
no heating
Instrument failure
Technical service
Paraffin dispenser
No paraffin flow
Paraffin not entirely
Check the consistency
molten
of the wax in the reser
voir
and wait for some time
if necessary
Foot pedal not properly
Check foot pedal
connected
connection
Whenever a malfunction occurs, please ensure that it is not the result of an operating mistake.
The list below includes the most frequent operating errors.
Equipment malfunctions are indicated as error messages (error codes 1 to 13) in the LCD display. If such an error
code is displayed, please contact your local Leica service center.
13.
Troubleshooting