Technical data – Leica Biosystems EG1160 User Manual

Page 6

Advertising
background image

6

Instruction manual V4.0 – 4/98

Power supply:

115 V / 60 Hz

230 V / 50 Hz

Fuses:
115V/60Hz

F1, F2 = T5.0 A

230V/50Hz

acc. to IEC 127-2

Power consumption:

800 VA

Operating
temperature range:

18°C - 35°C

Cooling system:
Refrigerant:

140 g ±5g R134A

Compressor oil:

150 ccm

EMKARATE RL15S

Dimensions:
(LxDxH)

910 x 575 x 320 mm

Weight:

50 kg

Capacities:

Paraffin reservoir:

3 l

Cassette bath:

approx.100 cass.

Heated
recessed area:

8 cassettes

Cold plate:

approx. 60 blocks

Programmable parameters:

Temperature

Paraffin reservoir
Mold warmer
Cassette bath
Work surface
Cold plate

Workdays
Daily work times (beginning, end)
Weekday
Time

Operation either manual or via the
foot pedal.

4.

Technical data

Functional unit

Temperature

Heated

Separate

Warm-up time

range

separately

temperature

prior to operation

control

Paraffin reservoir

45 - 70°C,

+

+

4 h

± 1°C increments

Mold warmer

35 - 70°C

+

+

1 h

± 1°C increments

Cassette bath

45 - 70°C

+

+

1.5 h

± 1°C increments

Work area

45 - 70°C

+

+

1 h

± 1°C increments

Cold plate

- 5°C

-

-

1 h

Forceps holder right

70°C

+

-

Forceps holder left

70°C

+

-

(not basic instrument)

Paraffin dispenser

45 - 70°C

+

-

+ pump

Temperature control

* at an ambient temperature of 22 °C and at an air humidity of 60%

All heated components include overheat protection.