Leica Biosystems EG1160 User Manual

Page 29

Advertising
background image

Leica EG 1160 – Paraffin Embedding Center

29

Paraffin dispenser

No paraffin flow

Foot pedal / dispenser

Technical service

valve switch failure

Paraffin drops when valve

Valve / pump failures

Technical service

closed

No heating

Heater (paraffin reservoir)

Activate heater (paraffin

not activated

reservoir)

Heater failure

Technical service

No illumination

Lamp defective

Replace lamp

Inhomogeneous paraffin

Air bubbles in dispenser

Push dispenser handle

flow (e.g. air bubbles)

at the start

for a while until flow is
homogeneous

Pump failure

Technical service

Mold warmer

Wrong temperature

Wrong temperature

Check temperature

Cassette bath

indication

selected

setting

Work area

No heating

Heater not activated

Activate heater

Heater failure

Technical service

Vacuum attachment

No vacuum buildup.

Lid not properly

Check if lid is properly

tional accessory)

Vacuum deteriorates too

positioned

(oppositioned

rapidly

Seal of lid contaminated

Check seal

Aeration knob open

Check knob
adjustment

Seal untight

Technical service

Forceps holder

No heating

Heater failure

Technical service

FUNCTIONAL UNIT

SYMPTOM

CAUSE

ACTION

13.

Troubleshooting