Leica Biosystems EG1160 User Manual
Page 30
Advertising
30
Instruction manual V 4.1 – 05/2001
Cold plate
Insufficient cooling
Ambient temperature too
Check temperature
high?
at lower ambient
temperature
Leak in cooling system?
Technical service
Air supply to compressor
Clean condenser fins
insufficient?
(Technical service)
Compressor does not
Compressor failure
Technical service
start
Compressor stops after
Compressor failure
Technical service
short operation
Compressor stops after
Compressor failure
Technical service
short operation and
restarts after a short while
FUNCTIONAL UNIT
SYMPTOM
CAUSE
ACTION
13.
Troubleshooting
Telephone number of your local
sales office:
Advertising