Operation – Leica Biosystems EG1160 User Manual

Page 22

Advertising
background image

22

Instruction manual V 4.1 – 05/2001

SET TEMPERATURES
PAR TANK

+XX°C

SET TEMPERATURES
PAR TANK

+XX°C

SET Mode (Programming)

6.4 WORKDAYS

The workdays are selected with the ARROW buttons. Releasing the but-
ton, the indicated day will be put in the memory automatically. The ON/
OFF button is used to confirm if the program shall be activated (YES) or
not (NO) the day indicated in the display.

If the programming of workdays is not desired, i.e. if the user can do
entirely without the automatic turn-on function of the tissue embedding
center, "NO" must be entered for each day of the week.

WORKDAYS
MO

YES

WORKDAYS
M

O

NO

6.5 How to quit the programming (SET) mode

1. 10 seconds after activation of any of the buttons, the in-

strument will automatically return to the ACT mode (de-
fault reading).
This feature is a protection against unintended program-
ming.

2. Pushing the SET/ACT button will also lead to the ACT

mode.

10 sec

The flow rate cannot be selected in the SET mode. It can only be
selected in the ACT mode.

AUTOMATIC TURN-ON FEATURE
If this feature is not used, it should be ensured that the tissue embed-
ding center is turned on early enough to allow all components to
reach the appropriate temperatures to avoid unnecessary delays.
Especially the paraffin reservoir heater should be turned on early
enough, as it can take some hours until the wax is molten. The indi-
vidual warmup times are listed in chapter 4 "Technical data".

10.

Operation

PAR TANK

+XX°C

TU

15:11

PAR TANK

+XX°C

TU

15:11

Advertising