Leica Biosystems EG1160 User Manual

Page 27

Advertising
background image

Leica EG1160 – Paraffin Embedding Center

27

2. Cleaning of paraffin reservoir, filter and dispenser outlet

The grid separating solid from liquid wax and the filter can be removed
from the paraffin reservoir for cleaning. The reservoir is cleaned inside
with a paper tissue.
Care should be taken that the reservoir is contaminated. Any dirt inside
should be removed prior to removing the grid and the filter.

3. Cleaning the forceps holder

The forceps holder, in particular, is frequently a source of contamination
and is susceptible to dirt. Therefore it should be cleaned thoroughly.

Caution:
The forceps holder is heated separately and thus very hot (approx. 70°C
or 80°C) during operation.

4. The paraffin drained to the paraffin collection tray should be emptied
regularly to ensure that excess paraffin can drain to the tray and prevent
the removable tray from sticking to guide.

5. Illumination of the dispenser outlet

Prior to exchanging the light bulb, the Leica EG 1160 must be turned off
with the green main switch on the back. Remove the cap at the front end
of the dispenser. The light bulb can be removed pulling it toward the user.
The replacement bulb is inserted by lightly pushing it between the con-
tacts.

6. Fuses

See "Technical data". Replacement fuses must meet the specifications
of the manufacturer. Otherwise, this could invalidate the warranty!

12.

Cleaning, maintenance, service

Advertising