Installation, 1 site requirements, 2 connection – Leica Biosystems ST5010 Autostainer XL User Manual

Page 14: 1 power, Site requirements, Connection

Advertising
background image

Instructions for Use V 2.3 – 10/2012

14

Instructions on how to install the instrument are provided in this chapter. A
diagram and description of components is also given. Finally, the procedure
for replacement of the fume filter is outlined.

4.1

Site requirements

The AutoStainer XL requires a solid bench of dimensions 1.090 mm long and
670 mm deep. The instrument must be located within 3 meters of a tap and
drain.

The power requirements for the AutoStainer XL are:

8 amps: at 110 volts
4 amps: at 240 volts

The voltage selector and other internal components are set by the manu-
facturer to suit the country of sale.

The voltage selector setting must not be altered by the user.

The AutoStainer XL requires connection to a laboratory water tap with mains
pressure fitting.

4.2

Connection

4.2.1 Power

Connect the power cord to the mains outlet ((28) page 11).

Connect the power jumper cable to the power supply outlet ((30) page
11) and to the power supply inlet ((32) page 11).

Refer to Figure 10 on page 11.

4.

Installation

Advertising