Leica AutoStainer XL – Automated slide stainer

6.1

Cleaning the instrument

Clean interior stainless steel surfaces with detergent and rinse with water.
Clean the head covers by wiping with a damp cloth.

The head contains sensitive electronic components. Do not
use liquids directly on this region. Wipe clean only.

The drain system may be flushed with 5% Sodium Hypochlorite to inhibit
microbial growth. If used, ensure that this solution does not remain in con-
tact with any metal parts for prolonged periods and flush well with water
after use. Exterior (painted) surfaces can be cleaned with a mild detergent
and wiped with a damp cloth.

Avoid the use of solvents on exterior surfaces and especially
on the control panel and lid.

Wipe the control panel carefully with a damp cloth.

6.1.1 Wash containers

Remove the wash containers and clean with detergent.

6.1.2 Reagent containers

Wash in warm water with detergent.

Do not wash reagent or wash containers in an automatic
dishwasher.

6.1.3 Slide racks

Clean with detergent or laboratory cleaning agent as required.

6.1.4 Oven

Periodically check the wax tray at the bottom of the oven and clean it if
excessive wax dripping has occurred.