Instructions for Use V 2.3 – 10/2012

16

4.3

Battery backup - UPS (optional)

An uninterruptable power supply (UPS) can be used to permit staining of
slides to continue during brief mains power failures.

A small UPS can be connected using the power jumper lead, as shown in
Figure 15.

The UPS should be rated at 200 VA for 5 minutes. Heating in the oven will not
be maintained by the UPS.

The UPS must be rated for use with the local mains voltage. Your distributor
can recommend a suitable UPS.