Leica AutoStainer XL – Automated slide stainer

5.4.8 Viewing a program

To view a program:

1. Select the program using steps 1 and 2, page 22.

2. Press [F3] View.

Up to four steps can be viewed simultaneously. Use the and

keys to view other steps.

3. Press [F4] to return to the previous screen.

5.4.9 Checking program compatibility

This function is used to check whether two programs can be run simulta-
neously. Programs cannot be run together if they need access to the same
exact station at the same time or if they contain the same two stations but in
reverse order, as in the following two programs:

Program 1

Program 2

Station 1

Station 1

Station 2

Station 3

Station 3

Station 2

1. Select the program (see steps 1 and 2, page 22).

2. Press [F2] Edit.

3. Press [F1] More.

4. Press [F2] Check.

5. Using the

and

keys, select the program number that you

wish to check compatibility with.

6. Press [F2] Check. - The compatibility check is performed and a subse-

quent

message will inform you whether the programs are compat-

ible.

7. If the programs are not compatible, an explanation will be given.

Press [F4] to continue.

8. Repeat Steps 5 and 6 to check compatibility with other programs.