Round Thick Diameter Specimens

Series 609 Alignment Fixture Product Information

Specimen Preparation

31

Round Thick Diameter Specimens

A typical gaged specimen has three levels of strain gages. The gages allow you to
determine bending strains at each level. Analysis of the strain readings produce a
profile of the grip alignment.

Gage placement

Use twelve gages on large diameter round specimens. Attach and number them
as shown here.

Gage Placement—Four Gage Round Specimens

Place the gages at the top, middle, and bottom of the gage section. Space
them at 90° intervals.

Place the top set of gages so that their upper edges just touch the upper edge
of the gage section.

Center the middle gages in the middle of the gage section.

Place the bottom set of gages so that their lower edges just touch the lower
edge of the gage section.

1

2

3

4

5

6

7

8

9

10

11

12

90°

180°

270°