Series 609 Alignment Fixture Product Information

66

Adjust the Angular Alignment

Alignment Procedure

The following figure shows typical readings before and after an adjustment.

Thin Flat Specimens—Reduce the “C” Bend

Thin flat specimens with three gage levels, four gages per level, turn the
angularity adjustment screws to try to get equal readings from all four middle
level gages.

If the magnitude of the average of the A side’s middle gage readings change, you
are overadjusting the angularity screws.

For example, if your original A side middle gage average goes from +10.5 to -
10.5, you are overadjusting.

Adjust

Strain readings

s (typical readings before adjustment)

t (typical readings after adjustment)

Try for zero
readings

Upper

Middle

Lower

Upper

Middle

Lower

Close to zero