Calculating Bending Strain—Notched Round Thin

Series 609 Alignment Fixture Product Information

Specimen Preparation

45

Gage Placement—Notched Three Gage Round Specimens

Place the gages at the top, middle, and bottom of the gage section. Space
them at 120° intervals.

Place the top set of gages so that their upper edges just touch the upper edge
of the gage section.

Place the bottom set of gages so that their lower edges just touch the bottom
of the gage section.

Calculating Bending Strain—Notched Round Thin Diameter Specimens

This section shows how to calculate the bending strain for round specimens with
three gages at each level.

Gage Placement

1

2

3

4

5

6

120°

240°