Flat Thick Specimens

Series 609 Alignment Fixture Product Information

Specimen Preparation

37

Flat Thick Specimens

A typical gaged specimen has three levels of strain gages. The gages allow you to
determine bending strains at each level. Analysis of the strain readings produces
a profile of the grip alignment.

Gage placement

Use twelve gages on thick flat specimens. Attach and number them as shown
here.

Gage Placement—Thick Flat Specimens

Place the gages in sets of four, at the top, middle, and bottom of the gage
section. Center each gage between the edges of the specimen.

Place the top set of gages so that their upper edges just touch the upper edge
of the gage section.

Place the middle set of gages in the middle of the gage section.

Place the bottom set of gages so that their lower edges just touch the lower
edge of the gage section.

1

2

3

4

5

6

7

8

9

10

11

12

90°

180°

270°