Flat thick specimens, Flat thick specimens 37, Gage placement – MTS Series 609 Alignment Fixture User Manual

Page 37

Advertising
background image

Flat Thick Specimens

Series 609 Alignment Fixture Product Information

Specimen Preparation

37

Flat Thick Specimens

A typical gaged specimen has three levels of strain gages. The gages allow you to
determine bending strains at each level. Analysis of the strain readings produces
a profile of the grip alignment.

Gage placement

Use twelve gages on thick flat specimens. Attach and number them as shown
here.

Gage Placement—Thick Flat Specimens

•

Place the gages in sets of four, at the top, middle, and bottom of the gage
section. Center each gage between the edges of the specimen.

•

Place the top set of gages so that their upper edges just touch the upper edge
of the gage section.

•

Place the middle set of gages in the middle of the gage section.

•

Place the bottom set of gages so that their lower edges just touch the lower
edge of the gage section.

1

2

3

4

5

6

7

8

9

10

11

12

0°

90°

180°

270°