Alignment Procedure

55

Alignment Procedure

This section covers alignment using the two most common gaged specimens.
They are the round specimen with three gage levels, four gages per level and the
thin flat specimen, also with three gage levels, four gages per level.

Note

It is important to start with the force transducer and actuator aligned. The
Series 609 Alignment Fixture is a precision instrument not intended to
correct gross misalignments. Using it to correct gross misalignments can
put high stresses on its mounting stud and shorten this stud’s fatigue life.

Note

Once you have completed the alignment, do not move the crosshead. If
you do, your alignment may change and you may have to repeat the
entire alignment procedure.

If you have a question about alignment using other gaged specimens, contact
MTS Systems Corporation (see the “Contact Information“ on the back of the title
page).

Contents

Installing a Gaged Specimen 56

Adjust the Concentric Alignment 58

Adjust the Angular Alignment 63

Check the Alignment 67

Prerequisite

You must have the grips installed in the load unit.