Adjust the Angular Alignment

Series 609 Alignment Fixture Product Information

Alignment Procedure

65

Round Specimens—Reduce the “C” Bend

If the magnitude of the middle gages’ readings change, you are overadjusting the
angularity screws.

Keep adjusting the screws to bring the readings closer together. Stop when your
readings start to move apart.

For example, if your original middle 180° reading goes from -15 to +15, you
have overadjusted.

Adjust

Strain readings

s (typical

t (typical

Try for zero
readings

Upper

Middle

Lower

Upper

Middle

Lower

Close to zero