Series 609 Alignment Fixture Product Information

48

Notched Flat Thick Specimens

Specimen Preparation

Notched Flat Thick Specimens

A typical gaged specimen has two levels of strain gages. The gages allow you to
determine bending strains at each level. Analysis of the strain readings produces
a profile of the grip alignment.

Gage placement

Use eight gages on thick flat specimens. Attach and number them as shown here.

Gage Placement—Notched Thick Flat Specimens

•

Place the gages in sets of four, at the top and bottom of the gage section.
Center each gage between the edges of the specimen.

•

Place the top set of gages so that their upper edges just touch the upper edge
of the gage section.

•

Place the bottom set of gages so that their lower edges just touch the lower
edge of the gage section.

4

3

0°

90°

180°

270°

6

5

1 2

7 8