Notched Flat Thick Specimens

Specimen Preparation

Notched Flat Thick Specimens

A typical gaged specimen has two levels of strain gages. The gages allow you to
determine bending strains at each level. Analysis of the strain readings produces
a profile of the grip alignment.

Gage placement

Use eight gages on thick flat specimens. Attach and number them as shown here.

Gage Placement—Notched Thick Flat Specimens

Place the gages in sets of four, at the top and bottom of the gage section.
Center each gage between the edges of the specimen.

Place the top set of gages so that their upper edges just touch the upper edge
of the gage section.

Place the bottom set of gages so that their lower edges just touch the lower
edge of the gage section.

4

3

90°

180°

270°

6

5

1 2

7 8