505.20/.30 HPU Connections

Model 505G2.20 - 505G2.30 SilentFlo™ HPU

Installation

27

Installation

Contents

505.20/.30 HPU Connections 27

Position the 505.20/.30 HPU 30

Testing the 505.20/.30 HPU 30

Precharging the Surge Suppressor Accumulator 35

505.20/.30 HPU Connections

Connecting the water

lines

For the purpose of cooling the hydraulic fluid, you must connect the HPU to a
water supply capable of maintaining the water flow rate as listed in the following
table.

The water supply must also be capable of maintaining water flow at a rate listed
in the following table.

The water supply must be equipped with an appropriate shutoff valve.

The differential pressure required between the HPU water inlet and outlet
connections is 0.24–0.34 MPa (35–50 psi).

The maximum allowable water inlet pressure is 0.83 MPa (120 psi).

Note

For hydraulic connections to air-cooled units, refer to the Air-Cooler to
SilentFlo HPU Integration Product Information manual (MTS part number
100-135-073).

To Connect the HPU to a

water supply

1. Connect the water supply to the HPU Water In port.

2. Connect the Water Out port to your drain/return line.

3. Adjust the water regulating valve to obtain the flow required by your water

temperature.

Water Inlet Temperature

Model 505.20

Model 505.30

15.5°C (60°F)

23.4 L/m (6.26 gpm)

35.2 L/m (9.3 gpm)

21.1°C (70°F)

30.2 L/m (8.06 gpm)

45.4 L/m (12 gpm)

26.7°C (80°F)

42.7 L/m (11.3 gpm)

64.7 L/m (17.1 gpm

32.2°C (90°F)

60.4 L/m (16 gpm)

91.2 L/m (24 gpm)