The “allow exclusive control at any level” control – MTS Model 494-05 Handset User Manual
Page 14
![background image](/manuals/568027/14/background.png)
MTS Model 494.05 Handset
Handset Setup Procedures
14
This is important because it prevents anyone from changing tuning and
calibration settings at the PC while the operator is using the handset to move the
actuator. While you do not use tuning and calibration controls to move the
actuator directly, they do affect actuator response.
The “Allow exclusive control at any level” Control
A control on the Handset Options page of the Station Options window (Station
Manager > Tools > Station Options > Handset) allows you to assign exclusive
control to the handset at any Station Manager application access level.
By default, this control is
disabled.
To enable this control, you
must attain the
Configuration access level
in the Station Manager
application.
Best practice for this control:
-Enable (check) this control only to facilitate tuning and calibration.
-Disable (click to clear) this control for normal operation.