Transducer, Transducer 72 – MTS SWIFT 45 GLP Sensor User Manual

Page 72

Advertising
background image

SWIFT

®

45 GLP Sensors

72

Maintenance

Transducer

The transducer requires a minimum amount of maintenance.

Do not pressure-wash the transducer or clean it with solvents.

Pressure-washing the transducer or cleaning it with solvents can damage it
or degrade the silastic seal and may void the warranty.

Using strong cleaners or solvents can damage the RTV seal and may void
the warranty.

Use only a soft sponge or brush with non-metal bristles and a gentle detergent
(such as dish soap) to wash the transducer.

Do not use high-pressure air to clean debris from around the transducer
connectors.

High-pressure air can damage the silastic seals and may void the warranty.

Use a brush with fine, non-metal bristles and low air-pressure [0.07 MPa (10 psi)]
to clean debris from around the transducer connectors.

As required

1. Inspect the transducer daily and/or after each testing session for any cracks

that may indicate fatigue, or physical damage that may indicate the
transducer has struck a hard surface or object. If the buildup of debris
prevents adequate inspection, perform Steps 2 and 3 as necessary, then
repeat this step.

Do not use a transducer if you determine that there are indications of fatigue
or other damage. Contact MTS.

2. Clean the transducer of debris after each testing session.

Carefully remove any debris (such as dust or gravel) that may be on the
transducer. If necessary use low-pressure-air (see the caution above).Do not
use a screwdriver or other rigid tool to pry out the debris. Prying may
damage the transducer.

3. Hand wash the transducer after each testing session (especially if the

transducer was exposed to corrosive and abrasive materials, such as salt or
sand) with a gentle detergent, such as dish soap, and a soft sponge or brush
with non-metal bristles. Be very careful when cleaning around areas where
RTV is used to seal the transducer so as not to break the seal.

4. Inspect the label affixed to the side of the transducer. Replace the label if it

becomes loose, has been lost or is unreadable.

CAUTION

CAUTION

Advertising