TI2SHUNT - Transducer Interface Shunt

This program is a utility with various functions related to shunts. The SWIFT
system includes the ability to connect a shunt resistor across each of the resistive
bridges in the transducer. This shunt function can be used as a simple verification
that the SWIFT system is working normally. Shunt verification activates the
shunts and compares the results to those recorded during calibration. While this
does not guarantee the transducer is still in calibration, it provides some level of
confidence it is working normally. If the shunt results differ significantly from
those recorded during calibration the SWIFT system should be evaluated for
possible problems.

Syntax

ti2shunt

Example

The Shunt Main Menu options are described in the following paragraphs.

Option 0

Use this option to exit the program.

Option 1

Use this option to read the last measured shunt values, the reference values, and
the shunt status.

Note

The shunt status is not maintained over power cycles, so it is only valid if
the shunt is executed after power is applied. Refer to

, “Error Messages,”

on page 38.

The following is typical of what is displayed when this selection is made:

C:\bin>ti2shunt

SWIFT Mini TI Shunt (Version 1.0)

0...Exit
1...Read current shunt status
2...Set the TI shunt tolerance
3...Scan inputs with shunts
4...Command a shunt cal
5...Set references to last measured

Enter choice: 1

Enter choice: 1

FX1 Ref:

0.854

Measured:

0.854

Status:

Good

FX2 Ref:

0.855

Measured:

0.855

Status:

Good

FY1 Ref:

0.857

Measured:

0.857

Status:

Good

FY2 Ref:

0.858

Measured:

0.857

Status:

Good

FY3 Ref:

0.856

Measured:

0.856

Status:

Good

FY4 Ref:

0.857

Measured:

0.856

Status:

Good

FZ1 Ref:

0.856

Measured:

0.856

Status:

Good

FZ2 Ref:

0.856

Measured:

0.855

Status:

Good