Download the Calibration File

Transducer Interface Setup

Download the Calibration File

Use the program TI2XFER to download the modified calibration file to the TI.

1. Insert the CD or diskette into the laptop computer or PC.

2. Run the program TI2XFER.

3. Enter 2 at the prompt to download the calibration file.

4. Enter the name of the file you wish to download.

5. TI2XFER will prompt you when the file has successfully downloaded.

6. Enter 0 at the prompt to exit the program.

SWIFT Mini TI transfer (Version 1.0)

Upload and download settings

0...Exit

1...Upload settings from TI box

2...Download settings to TI box

Choice? 2

Filename? sample.cal

0...Exit

1...Upload settings from TI box

2...Download settings to TI box

Choice? 0

Program completed