Transport, scope of delivery, installation

23

CAUTION

Risk of injury and danger of frostbite
Bursting of the inlet hose.

Depending on the operating state of the device, liquid nitrogen may be
enclosed in the supply line. When heated this may lead to the inlet hose
bursting.

The maximum pressure in the external liquid nitrogen supply line

may be a maximum of 1.5 bar. The use of a safety valve is absolutely
essential!

The connection for the liquid nitrogen (LN2) is a G 1/4 inch threaded connector .

The operating pressure in the liquid nitrogen supply line should be at least 0.5 bar.
The lower the pressure of the liquid nitrogen supply line, the longer the precooling
takes. At 0.5 bar supply line pressure and with the system at room temperature,
precooling requires approx. 10 minutes. Depending on the application, you can
execute up to 4 grinding operations with 10 litres of liquid nitrogen.

Throughout the entire duration of cooling, the air humidity in the chamber
condenses on all cooled parts as a layer of ice .

After cooling this condensation drips into the collecting filter and evaporates.

Pos : 5. 35 /00004 Warnhinweis e/ V0070 VOR SICHT LN 2 Austritt (CryoMill) @ 4\ mod_1314082103623_9. doc @ 25491 @ @ 1

CAUTION

Risk of injury and danger of frostbite
Uncontrolled escape of liquid nitrogen

There is a risk of injury in the event of an uncontrolled escape of liquid
nitrogen.

Stop the liquid nitrogen supply line immediately!

Pos : 5. 36 /00004 Warnhinweis e/H 0063 HIN WEIS Was ser aus dem Auff angfilter tropf en (CryoMill) @ 4\ mod_1314082101389_9. doc @ 25464 @ @ 1

NOTICE

Depending on the ambient temperature, air humidity and duration of use, small
quantities of water may drip from the collecting filter.

Pos : 5. 37 /00005 Ü berschrif ten/ 1. 1 Ü bersc hrift en/ 1. 1 Übersc hrift en BD A/ 11 Kühlmit telz ufl uss entf ernen @ 1 \ mod_1238053128286_9. doc @ 7970 @ @ 1

4.10 Removing the liquid nitrogen feed line

Pos : 5. 38 /00010 Bedi enungsanleit ung en Kapitels ammlungen/Cr yoMill II /0006 Tr ansport, Li ef erumf ang und Aufs tell en / 0012 Cr yoMill II M odul Kühl mitt elz ufluss entf ernen @ 5\ mod_1344511338025_9. doc @ 33963 @ @ 1

Abb. 13: Removing the liquid nitrogen feed line

CAUTION

Before you remove the liquid nitrogen feed line, the pressure must be relieved for
safety reasons.