5.4 Operating elements and displays

Pos : 6. 12 /00010 Bedi enungsanleit ung en Kapitels ammlungen/Cr yoMill II /0007 Bedi enung/0002 Cr yoMill II Modul Anz eige und Bedi en einheit @ 5\ mod_1343829230960_9.doc @ 33158 @ @ 1

Fig. 19: Control unit

Element

Description

Function

B1

LED on

– valve open

LED off

– valve closed

Indicates whether the internal solenoid valve on the
liquid nitrogen supply line is open or closed.

B2

LED on

– liquid nitrogen is flowing

LED off

– no liquid nitrogen

Indicates that liquid nitrogen is flowing into the device.

B3

LED precooling is running (Vkz)

Precooling is taking place / lights up when values are
set

B4

LED grinding is running (Mz)

The sample is ground / lights up when values are set

B5

LED intermediate cooling is running
(Zkz)

Intermediate cooling is taking plate / lights up when
values are set

B7

Frequency minus button

In setting mode and in operating mode:
reduces the set frequency by 1 Hz (1/s). Continuous
pressing switches on fast scrolling of numbers.

Frequency plus button

In setting mode and in operating mode:
increases the set frequency by 1 Hz (1/s). Continuous
pressing switches on fast scrolling of numbers.

B8

Cooling cycle minus button

Reduces the number of total cycles by 1

Cooling cycle plus button

Increases the number of total cycles by 1

B9

Time minus button

In setting mode and in operating mode:
Reducing the set

Precooling

Grinding time

Intermediate cooling time

B10