Operating the machine

36

Fig. 27: Inserting the grinding jar adapter

Pos : 6. 25 /00005 Ü berschrif ten/ 1. 1. 1 Ü bersc hrift en/ 111 Adapter f ür 4/2 x 2 ml Eppendorf Reaktionsgef äß e eins et zen @ 1 \ mod_1243499921893_9. doc @ 10328 @ @ 1

5.5.4

Insert the adapter for 6 / 4 / 2 x 2-ml Eppendorf reaction vial

Pos : 6. 26 /00010 Bedi enungsanleit ung en Kapitels ammlungen/Cr yoMill/0101 Warnhinw eis e/Cr yoMill HIN WEIS Zer brec hen von Eppendor f R eaktionsgef äß en @ 1 \mod_1243499975784_9. doc @ 10336 @ @ 1

NOTICE

Breaking of Eppendorf reaction vials
Changing material properties

The Eppendorf reaction vials that are available as accessories change
their material properties during the cryogen grinding.

Do not reuse reaction vials after the cryogen grinding.

Do not insert reaction vials in centrifuges after the cryogen grinding

Grinding may take place with a maximum 25Hz.

Pos : 6. 27 /00010 Bedi enungsanleit ung en Kapitels ammlungen/Cr yoMill II /0007 Bedi enung/0008 Cr yoMill II Modul Eppendorf G ef äße ei ns etz en @ 5\ mod_1343829240370_9. doc @ 33224 @ @ 1

You can operate the CryoMill with 2, 4 or 6 Eppendorf reaction vials (2ml).

NOTICE

Grinding with Eppendorf reaction vials may only take place with a maximum 25Hz.
(frequency 1/s)

Insert the 2ml Eppendorf reaction vials into the adapter for reaction vials as
shown in the diagram below.

Fig. 28: Inserting 2ml Eppendorf reaction vials