25 • Setup Mode

Eagle Instruction Manual

• To exit setup mode, from the main menu place the prompt next to

the last menu option, START MEASUREMENT, then press the
POWER/ENTER button. The Eagle begins its normal start-up
sequence.

Entering Setup Mode

WARNING: The Eagle does not detect gas or display readings while

in setup mode. The CAL/SETUP switch (SW2) must be in
the ON position to enter setup mode.

1. Take the Eagle to a non-hazardous location, and turn the power

off.

2. Press and hold down the AIR/▲ and SHIFT/▼ buttons, then

press the POWER/ENTER button. The main menu displays. It
displays four menu options at a time. Press the AIR/▲ or SHIFT/
▼ button to view additional menu options.

Updating the Battery Type Setting

This setting allows you to select between alkaline and Ni-Cd batteries.
The Eagle uses this setting to make sure adequate time is given
between the low battery warning and low battery alarm. This setting
has no effect on battery charging.

1. From the main menu, select the BATTERY TYPE menu option.

2. Press the AIR/▲ or SHIFT/▼ button to display the desired

setting.

CAUTION:

This setting should always match the type of batteries (alkaline
or Ni-Cd) installed in the Eagle. If this setting does not match
the installed batteries, the time between low battery warning
and low battery alarm may be less than expected.

3. Press the POWER/ENTER button to enter the setting and return to

the main menu.