Eagle Instruction Manual

Setup Mode • 28

setting. The fullscale setting flashes.

The maximum fullscale setting for the combustible gas channel is
50,000 PPM; the minimum setting is 1000 ppm. The default setting
is 50,000 ppm.

2. Press the AIR/▲ and SHIFT/▼ buttons to display the desired

fullscale setting (see Table 5), then press the POWER/ENTER
button to enter the setting. The prompt flashes.

Returning to the Main Menu

1. Press the SHIFT/▼ button. The

ESCAPE

message displays. (Press

the AIR/▲ button to return to the previous screen.)

2. Press the POWER/ENTER button to save the new setting. The

OTHER GAS SET

message displays, then the Gas Combinations

menu displays.

3. To exit the Gas Combinations menu, press the SHIFT/▼ button

until the prompt is next to Channel 4, then press SHIFT/▼ again.
The

ESCAPE

message displays.

4. Press the POWER/ENTER button. The message

SAVING DATA

displays, then the main menu displays.

Table 5: Fullscale PPM Readings

Equivalent to 100% LEL

Target Gas

Fullscale Setting

Methane (CH

4

)

50,000 ppm

Hexane

11,000 ppm

Hydrogen

40,000 ppm

Pentane

15,000 ppm

Styrene

9,000 ppm

IPA

20,000 ppm

Isobutane

18,000 ppm

Propane

21,000 ppm

Propylene

20,000 ppm

Toluene

11,000 ppm

Ethane

30,000 ppm

Ethanol

33,000 ppm

Benzene

12,000 ppm